4-62
Agilent 7500 ICP-MS ChemStation Operators Manual
Tuning
Tuning for Sensitivity (Agilent 7500cs)
Adjust sensitivity of elements to be measured obtain high stable signals over the
whole mass range. The standard tuning solution is a 1% nitric acid solution
containing 1 ng/mL (ppb) of each of the following: Li, Co, Y, Ce, and Tl.
When using the Agilent 7500a refer to “Tuning for Sensitivity (Agilent 7500a)” on
page 4-4.
When using the Agilent 7500ce, refer to “Tuning for Sensitivity (Agilent 7500ce)”
on page 4-39.
CAUTION Whenever the Agilent 7500cs is used, the ShieldTorch must be installed in it. For
installation of the ShieldTorch, refer to the Agilent 7500 ICP-MS Hardware Manual.
Adjust sensitivity using the following steps:
1 Select Instrument>>Tune in the ICP-MS Top Window.
The ICP-MS Tuning-Sensitivity Window will appear.
ICP-MS Tuning-Sensitivity Window