4-101
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Tuning
Loading a Tune File
To use a saved tune file for a tuning procedure, load the file by completing the
following steps in the Tune window:
1 Select File>>Load Tune Values.
The Load Tuning Parameter File dialog box appears.
Load Tuning Parameter File Dialog Box
2 Select the tune file in one of the following ways:
• Type the name of the tune file and click OK.
• Double-click a tune file in the displayed list.
• Click a tune file in the displayed list and click OK.
Agilent 7500 ChemStation will immediately change tuning parameters according to
the selected tune file.
NOTE If you change parameter values during the tuning procedure and want to save the new
values, save the tune file again. Otherwise, the changes are saved in atune.u only.
When you save, you can either overwrite or make a new tune file by renaming it.