5-63
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Creating a Method
Isotope Analysis [Multi Tune]
Isotope Analysis [Multi Tune] is similar to Spectrum Analysis [Multi Tune]. The
difference is that the dwell time per mass is 10 times shorter. This mode should be
used when precision of isotopic information (isotopic ratio and isotope dilution) is
required (e.g. radioactive isotope analysis for determining age, etc.).
To set the Acquisition Mode to Isotope Analysis [Multi Tune], complete the follow-
ing steps:
1 Click the Isotope Analysis [Multi Tune] button.
The button is marked and the selected mode is outlined.
Acquisition Mode Dialog Box---Isotope Analysis [Multi Tune]
2Click OK.
ChemStation sets Isotope Analysis as the acquisition mode for the method and the
Isotope Analysis [Multi Tune] Acquisition Parameters dialog box appears