5-53
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Creating a Method
Setup
This document will give you an example of the operation procedure (cool and hot)
from setting up the system to make a method using Multi Tune.
1 Rotary Pump
Confirm that only one Rotary Pump is running when running cool plasma.
Check the top on the valve between the two Rotary Pumps is rotated clockwise
(closed).
Check its breaker located behind the ICP-MS mainframe is turned off.
Turn off
Rotate
Clockwise