Section
3: Functions and features Model 2461 Interactive SourceMeter® Instrument
Reference Manual
3-46 2461-901-01 A/November 2015
Storing configuration index 3
Use one of the following methods to:
Set the time it takes to perform the current measurement (the NPLC setting) to 2.
Set the instrument source level to 4.0 V.
Store all active source settings to MySourceList as configuration index 3 by appending to the
end of the list.
In this example, the NPLC is set to 2 to demonstrate that a setting not part of the source configuration
list settings will not be saved.
Using SCPI commands:
:SENSe:CURRent:NPLCycles 2
:SOURce:VOLTage:LEVel 4
:SOURce:CONF:LIST:STORe "MySourceList"
Using TSP commands:
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.nplc = 2
smu.source.level = 4
smu.source.configlist.store("MySourceList")
The following figure shows the active state of the instrument after you change the source level to
4.0 V. Notice that the NPLC setting is not stored in MySourceList because this setting is not a
stored source setting.
Refer to Instrument settings stored in a source configuration list (on page 3-35) for a complete list of
source settings that the instrument stores in a source configuration list.
Figure 108: Example configuration point 3

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV