102                 Keysight E4980A/AL Precision LCR Meter 
Configuring Measurement Conditions (Display and Function Related Settings)
MEAS SETUP page
4-
Figure 4-5  Trigger delay time and step delay time
Table 4-5  Resolution steps when setting the trigger delay time with INCR + / DECR -
Trigger delay time Resolution
0 s through 100 ms 100 s
100 ms through 1 s 1 ms
1 s through 10 s 10 ms
10 s through 100 s 100 ms
100 s through 999 s 1 s
In DCR measurement, there are two step delay times because two cycles 
of measurement are performed by applying +0.1 V and -0.1 V. Accordingly, 
when measuring the Rdc parameter such as Ls-Rdc, since the stimulus 
setting and measurement procedure is executed three times in one 
measurement, the step delay time takes three times as long as the setting 
value.