Keysight E4980A/AL Precision LCR Meter 115
Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
Open Correction
The open correction feature of the E4980A/AL compensates for any stray
admittances (G, B) that may exist within the interval from the calibration plane,
which is determined by the selected cable length, to the DUT connecting
points (see Figure 4-9).
Figure 4-9 Stray admittances
Description of Correction Based on All Frequency Points
In this correction method, open correction data are collected from all 51
predetermined frequency points, regardless of the test frequency settings. At
the other frequency points, correction is done using data calculated by
interpolating the data collected at the 51 points (see Figure 4-10). The
following tables list the 51 predetermined frequency points:
Table 4-7 Predetermined frequency points in correction based on all frequency points
Test frequency points
20 Hz 100 Hz 1 kHz 10 kHz 100 kHz 1 MHz
25 Hz 120 Hz 1.2 kHz 12 kHz 120 kHz 1.2 MHz
30 Hz 150 Hz 1.5 kHz 15 kHz 150 kHz 1.5 MHz
40 Hz 200 Hz 2 kHz 20 kHz 200 kHz 2 MHz
50 Hz 250 Hz 2.5 kHz 25 kHz 250 kHz
60 Hz 300 Hz 3 kHz 30 kHz 300 kHz
80 Hz 400 Hz 4 kHz 40 kHz 400 kHz
500 Hz 5 kHz 50 kHz 500 kHz
600 Hz 6 kHz 60 kHz 600 kHz
800 Hz 8 kHz 80 kHz 800 kHz