- 448 Keysight E4980A/AL User’s Guide
Bias Current Interface
Product Overview
-
Product Overview
42841A Bias Current Source
The 42841A is a DC bias current source for use with the E4980A/AL (with
Option 002 Bias Current Interface). All measurement setups, including the DC
current bias, can be set from the E4980A/AL front panel.
Operation Limitations
Some E4980A/AL functions have the following limitations when used with the
42841A and the 42842A/B.
o Option 001, Internal DC Bias is disabled.
o DC Bias test signal voltage (V) is disabled.
o Test signal voltage monitor (VDC MON) is disabled.
o Test signal current monitor (IDC MON) is disabled.
o Auto bias polarity control function (BIAS POL) cannot be set to AUTO.
o Measurement range is only 10 and 100 .
o Cable length setting applies only to the 1 m setting.
Turn off all the power switches of the 42841A and E4980A/AL when
connecting or disconnecting any connection involving 42841A and
E4980A/AL.
The 42842C bias current test fixture cannot be used with the E4980A/AL.

Table of Contents

Other manuals for Keysight E4980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E4980A and is the answer not in the manual?

Keysight E4980A Specifications

General IconGeneral
Frequency Range20 Hz to 2 MHz
Basic Accuracy0.05%
DC Bias Voltage0 V to ±40 V
Measurement FunctionsL, C, R, Z, D, Q
DC Bias Current0 A to 20 mA (using external DC bias source)
DisplayLCD
InterfaceLAN, USB, GPIB
Dimensions426 mm (W) x 177 mm (H)
Weight13.5 kg