Keysight E4980A/AL Precision LCR Meter 81
Display Format
BIN COUNT DISPLAY Page
Counter Function
The E4980A/AL is capable of counting bins. The number of devices sorted into
each bin is counted while the unit sorts the devices into appropriate bins using
the comparator function.
The maximum count is 999,999. The counting operation stops and the overflow
message “----” appears when this value is reached.
Counter setting procedure
Step 1. Press the [Display Format] key.
Step 2. Press the BIN COUNT softkey.
Step 3. Use the cursor keys to select the COUNT field.
Step 4. Use the softkeys to set the counter to ON/OFF.
Counter reset procedure
Step 1. Press the [Display Format] key.
Step 2. Press the BIN COUNT softkey.
Step 3. Use the cursor keys to select the COUNT field.
Step 4. Use the softkeys to press RESET COUNT.
Softkey Function
COUNT ON Sets the counter to ON
COUNT OFF Sets the counter to OFF
The counter is reset when the instrument setting state is recalled.