348 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Equivalent key [Display Format] - MEAS DISPLAY - RANGE - AUTO
[Display Format] - MEAS DISPLAY - RANGE - HOLD
[Meas Setup] - MEAS SETUP - RANGE - AUTO
[Meas Setup] - MEAS SETUP - RANGE - HOLD
:FUNCtion:IMPedance:RANGe[:VALue]
Syntax :FUNCtion:IMPedance:RANGe[:VALue] <numeric>
:FUNCtion:IMPedance:RANGe[:VALue]?
Description Selects the impedance measurement range. This command turns the auto
range function OFF.
Parameter
Equivalent key [Display Format] - MEAS DISPLAY - RANGE
[Meas Setup] - MEAS SETUP - RANGE
:FUNCtion:IMPedance[:TYPE]
Syntax :FUNCtion:IMPedance[:TYPE]
{CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS|LPD|LPQ|LPG|LPRP|LPRD|LSD|LSQ|LSRS
|LSRD|RX|ZTD|ZTR|GB|YTD|YTR|VDID}
:FUNCtion:IMPedance[:TYPE]?
Description Selects the measurement function.
Parameter
<Numeric>
Range 100m|1|10|100|300|1k|3k|10k|30k|100k
Preset value 100
Unit ohms
Description
CPD(Preset value) Sets the Impedance Parameter Type to “Cp-D”
CPQ Sets the Impedance Parameter Type to “Cp-Q”
CPG Sets the Impedance Parameter Type to “Cp-G”
CPRP Sets the Impedance Parameter Type to “Cp-Rp”