338 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
viVQueryf(AgtE4980A/AL, "CORR:USE:DATA? 0" + vbLf, "%t",
Result)
Equivalent key [Meas Setup] - CORRECTION - MODE - MULTI
[Meas Setup] - CORRECTION - CH
[Meas Setup] - CORRECTION - OPEN A
[Meas Setup] - CORRECTION - OPEN B
[Meas Setup] - CORRECTION - SHORT A
[Meas Setup] - CORRECTION - SHORT B
[Meas Setup] - CORRECTION - LOAD A
[Meas Setup] - CORRECTION - LOAD B
:CORRection:USE:DATA:SINGle
Syntax :CORRection:USE:DATA:SINGle <open1 A>,<open1 B>,<short1 A>,<short1
B>,<load1 A>,<load1 B>,<open2 A>,<open2 B>,<short2 A>,<short2 B>,<load2
A>,<load2 B>,.......,<open201 A>,<open201 B>,<short201 A>,<short201
B>,<load201 A>,<load201 B>
:CORRection:USE:DATA:SINGle?
Description Sets or returns the open/short/load data for the “Single” correction mode. The
data for all 201 test must be set. Any point with no correction data should be
set to 0. A query will return 1,206 data for the 201 test points, even if there are
some frequency points that were turned off. A value of 0 is returned when there
are no correction data.
Parameter
Example of Use
Dim Result As String * 50000
viVQueryf(AgtE4980A/AL, "CORR:USE:DATA:SING?" + vbLf, "%t",
Result)
Equivalent key [Meas Setup] - CORRECTION - MODE - SINGLE
[Meas Setup] - CORRECTION - OPEN A
[Meas Setup] - CORRECTION - OPEN B
<open(n) A>, <open(n) B>, <short(n) A>, <short(n)
B>, <load(n)A>, <load(n) B>
Preset value 0
Unit OPEN: S; SHORT: ohms; LOAD: Depends on
:CORRection:LOAD:TYPE