136                 Keysight E4980A/AL Precision LCR Meter 
Configuring Measurement Conditions (Display and Function Related Settings)
LIMIT TABLE SETUP Page
4-
Turning On/Off the Comparator
Functional Description
The E4980A/AL’s built-in comparator can sort DUTs into a maximum 10 levels 
(bin 1 through bin 9 and OUT OF BINS) using up to nine sets of primary 
parameter limits along with one set of secondary parameter limits. 
In addition, DUTs that fall within the primary parameter limits but are out of the 
secondary parameter limits can be sorted into an auxiliary bin. 
To turn on or off the comparator: 
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Using the cursor keys, select the COMP field.
Step 4. Turn on or off the comparator by pressing the appropriate softkey: 
Softkey Description
ON Turns ON the comparator.
OFF Turns OFF the comparator.