:BIAS:CURRent[:LEVel]
Syntax :BIAS:CURRent[:LEVel] <numeric>
:BIAS:CURRent[:LEVel]?
Description Sets the DC bias current. Setting does not implicitly turn the DC bias ON. This
command is effective when option 001is installed. Without Option 001, this
command causes an error. If the bias is set by :BIAS:VOLTage[:LEVel], the
query command returns error No. 230. The parameter resolution depends on
the current setting value.
Parameter
Equivalent key [Display Format] - MEAS DISPLAY - BIAS
[Meas Setup] > - MEAS SETUP - BIAS
Description
SHORt Sets measurement time to Short
MEDium (Preset value) Sets measurement time to Medium
LONG Sets measurement time to Long
<Numeric>
Range 1 to 256
Preset value 1
Resolution 1
<Numeric>
Range
-100 m to 100 m
1
(with option 001)
Unit A