EasyManua.ls Logo

Keysight E4980A - Current[:Level]; Display:cclear; Display:enable

Keysight E4980A
519 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E4980A/AL Precision LCR Meter 339
SCPI Command Reference
E4980A/AL commands
[Meas Setup] - CORRECTION - SHORT A
[Meas Setup] - CORRECTION - SHORT B
[Meas Setup] - CORRECTION - LOAD A
[Meas Setup] - CORRECTION - LOAD B
:CURRent[:LEVel]
Syntax :CURRent[:LEVel] <numeric>
:CURRent[:LEVel]?
Description Sets the current level for the measurement signal. If the signal is set using
:VOLTage[:LEVel], the query command returns error No. 230.
Parameter
Equivalent key [Display Format] - MEAS DISPLAY - LEVEL
[Meas Setup] - MEAS SETUP - LEVEL
:DISPlay:CCLear
Syntax :DISPlay:CCLear
Description Clears errors or caution messages from the display. (No Query)
Equivalent key No equivalent key is available on the front panel.
:DISPlay:ENABle
Syntax :DISPlay:ENABle {ON|OFF|1|0}
:DISPlay:ENABle?
Description Enables display updates
<Numeric>
Range
0 to 100m
1
Unit A
Resolution 1
1. Depending on the DC Bias level.