EasyManua.ls Logo

Keysight E4980A - Function:smonitor:IDC[:State]; Function:smonitor:vac[:State]

Keysight E4980A
519 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
350 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Parameter
Equivalent key No equivalent key is available on the front panel.
:FUNCtion:SMONitor:IDC[:STATe]
Syntax :FUNCtion:SMONitor:IDC[:STATe] {ON|OFF|1|0}
:FUNCtion:SMONitor:IDC[:STATe]?
Description Enables the DC current-level monitor function.
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - IDC MON - ON
[Meas Setup] - MEAS SETUP - IDC MON - OFF
:FUNCtion:SMONitor:VAC[:STATe]
Syntax :FUNCtion:SMONitor:VAC[:STATe] {ON|OFF|1|0}
:FUNCtion:SMONitor:VAC[:STATe]?
Description Does nothing. The E4980A/AL always has its AC voltage-level monitor turned
ON. This function is provided because of its command compatibility with
4284A.
Parameter
Description
ON or 1 (Preset value) Does nothing
OFF or 0 Does nothing
This is available when option 001 is installed.
Description
ON or 1 DC current monitor function ON
OFF or 0 (Preset value) DC current monitor function OFF
Description
ON or 1 (Preset value) Does nothing