EasyManua.ls Logo

Keysight E4980A - Correction:spot[1-201]:Frequency; Correction:spot[1-201]:Load[:Execute]; Correction:spot[1-201]:Load:standard

Keysight E4980A
519 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E4980A/AL Precision LCR Meter 335
SCPI Command Reference
E4980A/AL commands
Equivalent key [Meas Setup] - CORRECTION - SHORT - ON
[Meas Setup] - CORRECTION - SHORT - OFF
:CORRection:SPOT[1-201]:FREQuency
Syntax :CORRection:SPOT[1-201]:FREQuency <numeric>
:CORRection:SPOT[1-201]:FREQuency?
Description Sets the frequency for the specified measurement point. The mode selected by
:CORRection:METHod is chosen as the correction mode for the specified point.
Parameter
Equivalent key [Meas Setup] - CORRECTION - FREQ
:CORRection:SPOT[1-201]:LOAD[:EXECute]
Syntax :CORRection:SPOT[1-201]:LOAD[:EXECute]
Description Executes LOAD correction at the specified measurement point. (No Query)
Equivalent key [Meas Setup] - CORRECTION - FREQ - MEAS LOAD
:CORRection:SPOT[1-201]:LOAD:STANdard
Syntax :CORRection:SPOT[1-201]:LOAD:STANdard <reference value for primary>,
<reference value for secondary>
:CORRection:SPOT[1-201]:LOAD:STANdard?
Description Sets the standard reference values at the specified measurement point. The
measurement function for the reference values can be selected using
:CORRection:LOAD:TYPE.
OFF or 0 (Preset value) Disables SHORT correction
Description
<Numeric>
Range 20 to 2M
Preset value 20
Unit Hz

Table of Contents

Other manuals for Keysight E4980A

Related product manuals