Keysight E4980A/AL Precision LCR Meter 375
SCPI Command Reference
E4980A/AL commands
:VOLTage[:LEVel]
Syntax :VOLTage[:LEVel] <numeric>
:VOLTage[:LEVel]?
Description Sets the voltage level for the measurement signal. If the signal is set by
:CURRent[:LEVel], the query command returns error No. 230.
Parameter
Equivalent key [Display Format] - MEAS DISPLAY - LEVEL
[Meas Setup] - MEAS SETUP - LEVEL
<Numeric>
Range
0 to 20
1
Preset value 1
Unit V
Resolution 100
1. Depending on the DC bias level.

Table of Contents

Other manuals for Keysight E4980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E4980A and is the answer not in the manual?

Keysight E4980A Specifications

General IconGeneral
Frequency Range20 Hz to 2 MHz
Basic Accuracy0.05%
DC Bias Voltage0 V to ±40 V
Measurement FunctionsL, C, R, Z, D, Q
DC Bias Current0 A to 20 mA (using external DC bias source)
DisplayLCD
InterfaceLAN, USB, GPIB
Dimensions426 mm (W) x 177 mm (H)
Weight13.5 kg