330 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
:COMParator:TOLerance:NOMinal
Syntax :COMParator:TOLerance:NOMinal <numeric>
:COMParator:TOLerance:NOMinal?
Description Sets the nominal value for the comparator function’s tolerance mode. This can
be set only when the limit mode is set to the tolerance mode.
Parameter
Equivalent key [Meas Setup] - LIMIT TABLE - NOM
:CONTrol:CBIas:STATe
Syntax :CONTrol:CBIas:STATe {ON|OFF|1|0}
:CONTrol:CBIas:STATe?
Description Enables the current bias I/F. If the 42841A is connected, the preset status is
ON. This is available only for serial prefix number below:
MY464/SG464 and below for E4980A
MY542/SG542 and below for E4980AL
Parameter
Equivalent key [System] - SYSTEM INFO - CURR BIAS I/F - ON
[System] - SYSTEM INFO - CURR BIAS I/F - OFF
:CONTrol:HANDler:STATe
Syntax :CONTrol:HANDler:STATe {ON|OFF|1|0}
:CONTrol:HANDler:STATe?
Description Enables the handler I/F.
<Numeric>
Preset value 0
Description
ON or 1 Enables the current bias I/F
OFF or 0 (Preset value) Disables the current bias I/F