362 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
[Meas Setup] - MEAS SETUP - DCI RNG - HOLD
:OUTPut:DC:ISOLation:LEVel:VALue
Syntax :OUTPut:DC:ISOLation:LEVel:VALue <numeric>
:OUTPut:DC:ISOLation:LEVel:VALue?
Description Selects DC isolation range. This command turns the auto range function OFF.
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - DCI RNG
:OUTPut:DC:ISOLation[:STATe]
Syntax :OUTPut:DC:ISOLation[:STATe] {ON|OFF|1|0}
:OUTPut:DC:ISOLation[:STATe]?
Description Enables DC Isolation.
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - DCI ISO - ON
[Meas Setup] - MEAS SETUP - DCI ISO - OFF
:OUTPut:HPOWer
Syntax :OUTPut:HPOWer {ON|OFF|1|0}
:OUTPut:HPOWer?
Description Does nothing. This function is provided because of its command compatibility
with 4284A.
<Numeric>
Range
20 |200 |2 m|20 m|100 m
Preset value 20 m
Unit A
Description
ON or 1 Enables DC Isolation
OFF or 0 (Preset value) Disables DC Isolation