Keysight E4980A/AL Precision LCR Meter 347
SCPI Command Reference
E4980A/AL commands
:FUNCtion:DEV[1-2]:REFerence:FILL
Syntax :FUNCtion:DEV[1-2]:REFerence:FILL
Description Executes a single measurement and enters two measured values (the primary
and secondary parameters) into each of the reference values for deviation
measurement. FUNC:DEV1:REF:FILL and FUNC:DEV2:REF:FILL are the
identical. (No Query)
The E4980A/AL operates in the same way with either DEV1 or DEV2.
Equivalent key [Meas Setup] - MEAS SETUP - REF A - MEASURE
[Meas Setup] - MEAS SETUP - REF B - MEASURE
:FUNCtion:DEV[1-2]:REFerence[:VALue]
Syntax :FUNCtion:DEV[1-2]:REFerence[:VALue] <numeric>
:FUNCtion:DEV[1-2]:REFerence[:VALue]?
Description Sets the reference value for deviation measurement.
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - REF A
[Meas Setup] - MEAS SETUP - REF B
:FUNCtion:IMPedance:RANGe:AUTO
Syntax :FUNCtion:IMPedance:RANGe:AUTO {ON|OFF|1|0}
:FUNCtion:IMPedance:RANGe:AUTO?
Description Enables the auto-range function for impedance measurement.
Parameter
<Numeric>
Preset value 0
Unit Depends on :FUNCtion:IMPedance[:TYPE].
Description
ON or 1 (Preset value) Auto range ON
OFF or 0 Auto range OFF