Keysight E4980A/AL Precision LCR Meter 181
Save/Recall
Saving Measurement Results into USB Memory
To save measurement results into USB memory:
For how to save the measurement result, refer to “How to save the
measurement result of List Sweep Measurement to USB memory.” on
page 182.
Step 1. Plug a USB memory stick into the front USB port.
Step 2. Press [Save/Recall].
Step 3. Press the SAVE DATA softkey.
Step 4. Press the START LOG softkey and then press the following softkeys to enter
the measurement results into the data buffer memory.
Step 5. Start measurement. The data buffer memory is filled with up to 201 sets of
measurement results.
Step 6. Press the SAVE & STOP softkey to save results into the USB memory.
Step 7. When the data has been saved into the USB memory, a “Storing data
completed. : E498xXXX.csv” message appears in the system message area.
Softkey Description
START LOG Starts logging the measurement results into the data buffer memory.
SAVE &
STOP
Copies the data from the data buffer memory into the USB memory.
Then stops saving the measurement results into the data buffer
memory and clears the data buffer memory.