Keysight E4980A/AL Precision LCR Meter 325
SCPI Command Reference
E4980A/AL commands
Parameter
Equivalent key [Display Format] - MEAS DISPLAY - BIAS
[Meas Setup] - MEAS SETUP - BIAS
:COMParator:ABIN
Syntax :COMParator:ABIN {ON|OFF|1|0}
:COMParator:ABIN?
Description Enables the auxiliary BIN counting function of the comparator.
Parameter
Equivalent key [Meas Setup] - LIMIT TABLE - AUX - ON
[Meas Setup] - LIMIT TABLE - AUX - OFF
:COMParator:BEEPer
Syntax :COMParator:BEEPer {FAIL|PASS}
:COMParator:BEEPer?
Description Selects the condition for generating a beep sound: when sorting with the
comparator fails (sorts into OUT OF BIN) or passes (sorts into BINs 1 to 9 or
AUX BIN). The beep should be turned on by the :SYSTem:BEEPer:STATe.
<Numeric>
Range 0,1.5,2 (without option 001)
-40 to 40
1
(with option 001)
Preset value 0
Unit V
1. Depending on signal level.
Description
ON or 1 Enables the auxiliary BIN counting function