Keysight E4980A/AL Precision LCR Meter 79
Display Format
BIN COUNT DISPLAY Page
BIN COUNT DISPLAY Page
When you press the [Display Format] key and BIN COUNT softkey, the BIN
COUNT DISPLAY page is displayed. On this page, the comparator’s count
results are displayed (The field in parentheses is used for setting).
• Counter function ON/OFF (COUNT field)
This page also provides the following information in the monitor areas (each
monitor area looks like a field, but is not). These conditions can be set from the
LIMIT TABLE SETUP page. For more details, refer to “LIMIT TABLE SETUP
Page” on page 130.
• Measurement Function (FUNC)
• Nominal Value (NOM)
• Test Frequency (FREQ)
• Bin Sorting Low/High Limits (LOW/HIGH)
• BIN Sorting Results (RESULT)
Figure 3-10 shows the available fields and the softkeys that correspond to
each field on this page.
You can set it so that a beep sounds when the sorting result is output. For
details, refer to “Beep Feature” on page 139.