EasyManua.ls Logo

Keysight E4980A - List:bias:voltage; List:clear:all; List:current

Keysight E4980A
519 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
354 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Equivalent key [Meas Setup] - LIST SETUP - BIAS[A]
:LIST:BIAS:VOLTage
Syntax :LIST:BIAS:VOLTage <point 1>,<point 2>,.... ,<point n>
:LIST:BIAS:VOLTage?
Description Clears the previous list sweep table, sets DC voltage sweep as the list sweep
parameter, and sets DC voltage points for the DC voltage list sweep. If the list
sweep is not set for DC voltage, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Equivalent key [Meas Setup] - LIST SETUP - BIAS[V]
:LIST:CLEar:ALL
Syntax :LIST:CLEar:ALL
Description Clears the list sweep setup. (No Query)
Equivalent key [Meas Setup] - LIST SETUP - No. - CLEAR TABLE
:LIST:CURRent
Syntax :LIST:CURRent <point 1>,<point 2>,... ,<point n>
:LIST:CURRent?
Description Clears the previous list sweep table, sets AC current sweep as the list sweep
parameter, and sets AC current points for the AC current list sweep. If the list
sweep is not set for AC current, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
<point n>
Range of n 1 to 201
Unit V
<point n>
Range of n 1 to 201
Unit A