EasyManua.ls Logo

Keysight E4980A - Fetch:smonitor:iac; Fetch:smonitor:IDC; Fetch:smonitor:vac; Fetch:smonitor:VDC

Keysight E4980A
519 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E4980A/AL Precision LCR Meter 343
SCPI Command Reference
E4980A/AL commands
Description Returns a measurement result with the selected measurement function. See
the “Read Measurement Results” on page 286. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:FETCh:SMONitor:IAC?
Syntax :FETCh:SMONitor:IAC?
Description Returns the latest measured data for the AC current monitor. In case of
Vdc-Idc measurement, 0 is returned. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:FETCh:SMONitor:IDC?
Syntax :FETCh:SMONitor:IDC?
Description Returns the latest measured data for the DC current monitor. If the
current-level monitor is set to OFF, 0.0 is returned. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:FETCh:SMONitor:VAC?
Syntax :FETCh:SMONitor:VAC?
Description Returns the latest measured data for the AC voltage monitor. In case of
Vdc-Idc measurement, 0 is returned. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:FETCh:SMONitor:VDC?
Syntax :FETCh:SMONitor:VDC?
Description Returns the latest measured data for the DC voltage monitor. If the
voltage-level monitor is set to OFF, 0.0 is returned. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:FORMat:ASCii:LONG
Syntax :FORMat:ASCii:LONG {ON|OFF|1|0}
:FORMat:ASCii:LONG?
Description Enables the long format (+0.0000000000E+00). This is applied to the returned
values of :FETCh[:IMPedance][:FORMatted]?/:MEMory:READ?/*TRG.

Table of Contents

Other manuals for Keysight E4980A

Related product manuals