EasyManua.ls Logo

Keysight E4980A - Correction:use[:Channel]; Correction:use:data[:Multi]

Keysight E4980A
519 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E4980A/AL Precision LCR Meter 337
SCPI Command Reference
E4980A/AL commands
:CORRection:USE[:CHANnel]
Syntax :CORRection:USE[:CHANnel] <channel>
:CORRection:USE[:CHANnel]?
Description Selects the channel number to be used for the Scanner I/F MULTI mode.
Parameter
Equivalent key [Meas Setup] - CORRECTION - CH
:CORRection:USE:DATA[:MULTi]
Syntax :CORRection:USE:DATA[:MULTi] <channel>,<open1 A>,<open1 B>,<short1
A>,<short1 B>,<load1 A>,<load1 B>,<open2 A>,<open2 B>,<short2 A>,<short2
B>,<load2 A>,<load2 B>,.......,<open 201A>,<open201 B>,<short201
A>,<short201 B>,<load201 A>,<load201 B>
:CORRection:USE:DATA[:MULTi]? <channel>
Description Sets or returns the open/short/load data for the “Multi” correction mode. The
data for all 201 test points must be set. Any point with no correction data
should be set to 0. A query will return 1,206 data for 201 test points even if
there are some frequency points that were turned off. A value of 0 is returned
when there are the no correction data.
Parameter
Example of Use
Dim Result As String * 50000
<channel>
Range 0 to 127
Preset value 0
<channel>
Range 0 to 127
<open(n) A>, <open(n) B>, <short(n) A>, <short(n)
B>, <load(n)A>, <load(n) B>
Preset value 0
Units OPEN: S; SHORT: ohms; LOAD: Depends on
:CORRection:LOAD:TYPE

Table of Contents

Other manuals for Keysight E4980A

Related product manuals