Keysight E4980A/AL Precision LCR Meter 353
SCPI Command Reference
E4980A/AL commands
Description Sets the limit values of the limit function for the list sweep measurement. When
A or B is the parameter, both lower and the upper limit should be input. If
either of them is not input, error No. 109 is returned. In the case of
OFF,<low>,<high> are optional.
Parameter
Equivalent key [Meas Setup] - LIST SETUP - LMT - A
[Meas Setup] - LIST SETUP - LMT - B
[Meas Setup] - LIST SETUP - LMT -
[Meas Setup] - LIST SETUP - LOW
[Meas Setup] - LIST SETUP - HIGH
:LIST:BIAS:CURRent
Syntax :LIST:BIAS:CURRent <point 1>,<point 2>,.... ,<point n>
:LIST:BIAS:CURRent?
Description Clears the previous list sweep table, sets DC current sweep as the list sweep
parameter, and sets DC current points for the DC current list sweep. If the list
sweep is not set for DC current, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Description
A Sets limit test target to A (Primary parameter)
B Sets limit test target to B (Secondary parameter)
OFF (Preset value) Sets limit test target to OFF
<low>,<high>
Preset value Query returns -9.9E37, +9.9E37 for <low>, <high>,
respectively in the case of OFF.
Unit Depends on :FUNCtion:IMPedance[:TYPE]
<point n>
Range of n 1 to 201
Unit A