214 Keysight E4980A/AL Precision LCR Meter
Measurement Procedure and Examples
Inductance Measurements
7-
Inductance Measurements
This section describes a practical example of measuring a magnetic-core
inductor.
The basic procedure flow to perform this measurement is the same as the Basic
Measurement Procedure described previously. A magnetic-core inductor is
measured under the following conditions.
Step 1. Turn the E4980A/AL ON.
Step 2. Set up the E4980A/AL measurement conditions by filling in the fields on the
MEAS DISPLAY page.
1. Move to the FUNC field using the cursor keys and select Ls-Rdc.
2. Move to the FREQ field using the cursor keys and input 100 kHz.
3. Move to the LEVEL field using the cursor keys and input 10 mA.
Step 3. Connect the test fixture to the E4980A/AL as shown in Figure 7-18. The
16047E Direct Couple Test Fixture (general purpose) is used for this
measurement.
Step 4. Perform the correction. To compensate for the Keysight 16047E’s residuals
and strays, an OPEN/SHORT correction is required.
1. Leave the 16047E in an OPEN condition as shown in Figure 7-18.
2. Press the [Meas Setup] key and the CORRECTION softkey.
3. Move to the OPEN field using the cursor keys and press the MEAS OPEN
softkey.
Wait until the message “OPEN measurement in progress” disappears.
4. Press the ON softkey to enable an OPEN correction.
5. Connect a shorting bar to the Keysight 16047E to set up the SHORT
condition as shown in Figure 7-19.
6. Move to the SHORT field using the cursor keys and press the MEAS SHORT
softkey.
Wait until the message “SHORT measurement in progress” disappears.
7. Press the ON softkey to enable a SHORT correction.
Sample
(DUT)
Magnetic-core inductor
Measurement
Conditions
Function: Ls-Rdc
Test frequency: 100 kHz
Test Signal Level: 10 mA (constant)

Table of Contents