360 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Equivalent key No equivalent key is available on the front panel.
:MEMory:FILL
Syntax :MEMory:FILL DBUF
Description Enables the data buffer memory to store measurement data. After execution of
this command, all measurement data will be stored in the data buffer memory.
If this has already been enabled, this command will do nothing. (No Query)
Equivalent key [Save/Recall] - SAVE DATA - START LOG
:MEMory:READ?
Syntax :MEMory:READ? DBUF
Description Places the data in the data buffer memory into the output buffer. The output
format of this command is the same as the one for FETC:IMP?. In the preset
state, “9.90000E+37,+9.90000E+37,-1,+0” is returned the number of times
specified by :MEMory:DIM. When the data buffer memory is not filled to the
specified size (specified by the :MEMory:DIM command),
“9.90000E+37,+9.90000E+37,-1,+0” remains for the rest of the data locations.
(Query Only)
Equivalent key [Save/Recall] - SAVE DATA - SAVE & STOP
This key is a combination of both :MEMory:READ? and :MEMory:CLEar.
:MMEMory:DELete[:REGister]
Syntax :MMEMory:DELete[:REGister] <numeric>
Description Deletes the state from the memories. Numbers 0 to 9 are located in the
internal memory, while Nos. 10 to 19 are situated in the external USB memory.
(No Query)
Parameter
Equivalent key [Save/Recall] - No. - DELETE
:MMEMory:LOAD:STATe[:REGister]
Syntax :MMEMory:LOAD:STATe[:REGister] <numeric>
<Numeric>
Range 0 to 19
Resolution 1

Table of Contents

Other manuals for Keysight E4980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E4980A and is the answer not in the manual?

Keysight E4980A Specifications

General IconGeneral
Frequency Range20 Hz to 2 MHz
Basic Accuracy0.05%
DC Bias Voltage0 V to ±40 V
Measurement FunctionsL, C, R, Z, D, Q
DC Bias Current0 A to 20 mA (using external DC bias source)