Keysight E4980A/AL Precision LCR Meter 163
System Configurations
SELF TEST Page
Choosing a Test Item
Functional Description
The E4980A/AL can run the following self tests:
To run a self test:
For step-by-step instructions, see “Daily Checks (Executing the self-test)” on
page 415.
Table 5-3
Test item Description
SYSTEM Checks the entire system, A1/A2/A3 boards, system
calibration data, and so on.
USER DATA Checks the GPIB/LAN configurations, instrument
configuration information, instrument calibration data, and
scanner calibration data.
BATTERY Checks the internal batteries.
KEY Checks the font panel keys. (Visual confirmation only; no
pass/fail result shown on screen.)
DISPLAY Checks the font panel LEDs and LCD. (Visual confirmation
only; no pass/fail result shown on screen.)
HANDLER
INTERFACE
Checks the handler interface. (Visual confirmation only; no
pass/fail result shown on screen.)
SCANNER
INTERFACE
Checks the scanner interface. (Visual confirmation only; no
pass/fail result shown on screen.)