364 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - DC SRC
:STATus:OPERation:CONDition?
Syntax :STATus:OPERation:CONDition?
Description Returns the value of the Operation Status Condition register. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:STATus:OPERation:ENABle
Syntax :STATus:OPERation:ENABle <numeric>
:STATus:OPERation:ENABle?
Description Sets the value of the Operation Status Enable register.
Parameter
Equivalent key No equivalent key is available on the front panel.
:STATus:OPERation[:EVENt]
Syntax :STATus:OPERation[:EVENt]
Description Returns the value of the Operation Status Event register. (Query Only)
Equivalent key No equivalent key is available on the front panel.
<Numeric>
Range -10 to 10
Preset value 0
Unit V
Resolution 1 m
<Numeric>
Range -32,768 to 32,767
Preset value 0