Keysight E4980A/AL Precision LCR Meter               357
SCPI Command Reference
E4980A/AL commands
:LIST:STIMulus:DATA
Syntax :LIST:STIMulus:DATA <Point 1 1stSTIM>,<Point 1 2ndSTIM>,<Point 2 
1stSTIM>,<Point 2 2ndSTIM>,....,<point n 1stSTIM>,<point n 2ndSTIM>
:LIST:STIMulus:DATA?
Description Sets the list sweep with two parameters (first and secondary parameters). This 
command allows you to sweep with two parameters. For example, you can 
sweep at point no.1 with 1 kHz/1 V, point no. 2 with 2 kHz/1.1 V, and point no. 
3 with 3 kHz/0.9 V. Both first and second parameters must be set unless the 
second parameter of :LIST:STIMulus:TYPE command is set to NONE. The value 
9.9E37 should be set for any empty points. When the IRANge is selected as the 
secondary parameter in the :LIST:STIMulus:TYPE command, the value for 
<Point n 2ndSTIM> can selected within the value of 
:FUNCtion:IMPedance:RANGe[:VALue].
Parameter
Example of Use
viVPrintf(AgtE4980A/AL, "LIST:STIM:TYPE FREQ, VOLT" + vbLf, 0)
viVPrintf(AgtE4980A/AL, "LIST:STIM:DATA 1E3, 1, 2E3, 1.1, 3E3, 
0.9" + vbLf, 0)
Equivalent key No equivalent key is available on the front panel.
:LIST:STIMulus:MDATa?
Syntax :LIST:STIMulus:MDATa?
Description Returns actual stimulus values for a list sweep with two parameters. This 
command allows you to know the actual applied values after the auto polarity 
function works. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:LIST:STIMulus:TYPE
Syntax :LIST:STIMulus:TYPE 
{FREQuency|VOLTage|CURRent|BVOLtage|BCURrent|DCSVoltage},{NONE|IRA
Nge|FREQuency|VOLTage|CURRent|BVOLtage|BCURrent|DCSVoltage}
<Point n 1stSTIM>,<Point n 2ndSTIM>
1
Range of n 1 to 201
Unit Depends on :LIST:STIMulus:TYPE
1. When the second parameter of :LIST:STIMulus:TYPEcommand is set to 
NONE, this parameter is not needed.