Keysight E4980A/AL Precision LCR Meter 361
SCPI Command Reference
E4980A/AL commands
Description Loads the state from the memories. Numbers 0 to 9 are located in the internal
memory, while Nos. 10 to 19 are situated in the external USB memory. (No
Query)
Parameter
Equivalent key [Save/Recall] - No. - RECALL
:MMEMory:STORe:STATe[:REGister]
Syntax :MMEMory:STORe:STATe[:REGister] <numeric>
Description Stores the state to the memories. Numbers 0 to 9 are located in the internal
memory, while Nos. 10 to 19 are situated in the external USB memory. Number
10 is automatically recalled. (No Query)
Parameter
Equivalent key [Save/Recall] - No. - SAVE
:OUTPut:DC:ISOLation:LEVel:AUTO
Syntax :OUTPut:DC:ISOLation:LEVel:AUTO {ON|OFF|1|0}
:OUTPut:DC:ISOLation:LEVel:AUTO?
Description Enables the auto range function for DCI.
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - DCI RNG - AUTO
<Numeric>
Range 0 to 19
Resolution 1
<Numeric>
Range 0 to 19
Resolution 1
Description
ON or 1 (Preset value) Enables the auto range function for DCI
OFF or 0 Disables the auto range function for DCI