Keysight E4980A/AL Precision LCR Meter 145
Configuring Measurement Conditions (Display and Function Related Settings)
LIST SWEEP SETUP Page
List Sweep Parameters
Functional Description
The sweep parameter used in list sweep measurement can be measurement
frequency, signal level, DC bias, or DC source. Use the sweep point field to
specify the list sweep measurement parameter.
To specify the list sweep measurement parameter:
Step 1. Press [Meas Setup].
Step 2. Press the LIST SETUP softkey.
Step 3. Using the cursor keys, select the sweep parameter field.
Step 4. Select your desired list sweep measurement parameter by pressing the
appropriate softkey:
Softkey Description
FREQ [Hz] Uses frequency as the list sweep parameter.
LEVEL [V] Uses voltage as the list sweep parameter.
LEVEL [A] Uses current as the list sweep parameter.
BIAS [V] Uses DC bias voltage as the list sweep parameter.
BIAS [A] Uses DC bias current as the list sweep parameter.
DC SRC [V] Uses DC source as the list sweep parameter.
When the sweep parameter is set using the SCPI commands, the
secondary sweep parameter can be set. In addition, the measurement
range can be set for the secondary parameter. For details, refer to
“:LIST:STIMulus:TYPE” on page 357.