Keysight E4980A/AL Precision LCR Meter 333
SCPI Command Reference
E4980A/AL commands
Equivalent key The softkeys of [Meas Setup] - CORRECTION - FUNC field
:CORRection:METHod
Syntax :CORRection:METHod {SINGle|MULTiple}
:CORRection:METHod?
Description Selects the correction mode (Single or Multi). Scanner I/F should be installed
for the Multi mode.
Parameter
Equivalent key [Meas Setup] - CORRECTION - MODE - SINGLE
[Meas Setup] - CORRECTION - MODE - MULTI
LPD Sets the type of LOAD correction reference to “Lp-D”
LPQ Sets the type of LOAD correction reference to “Lp-Q”
LPG Sets the type of LOAD correction reference to “Lp-G”
LPRP Sets the type of LOAD correction reference to “Lp-Rp”
LSD Sets the type of LOAD correction reference to “Ls-D”
LSQ Sets the type of LOAD correction reference to “Ls-Q”
LSRS Sets the type of LOAD correction reference to “Ls-Rs”
RX Sets the type of LOAD correction reference to “R-X”
ZTD Sets the type of LOAD correction reference to “Z-thd”
ZTR Sets the type of LOAD correction reference to “Z-thr”
GB Sets the type of LOAD correction reference to “G-B”
YTD Sets the type of LOAD correction reference to “Y-thd”
YTR Sets the type of LOAD correction reference to “Y-thr”
Description
Description
SINGle (Preset value) Sets the correction mode to “SINGLE”
MULTiple Sets the correction mode to “MULTI”