- 482 Keysight E4980A/AL User’s Guide
Handler Interface
Using the Handler Interface
-
Signals Activated when Overload Occurs
Signals activated when overload occurs are as follows. For conditions where
overload occurs, refer to “Displaying Errors instead of Measurement Results”
on page 71.
Table E-15 Signals activated when Overload Occurs
Measurement Parameter that overload occured Handler output
(active signal)
Impedance measurement (except for Ls/Lp-Rdc) /OUT_OF_BINS
/OVLD
Impedance measurement (Ls/Lp-Rdc) /OUT_OF_BINS
/OVLD
DCR measurement (Ls/Lp-Rdc) /OUT_OF_BINS
/OVLD
Vdc measurement (primary) /OUT_OF_BINS
/OVLD
Idc measurement (primary) /OUT_OF_BINS
/OVLD
Vdc measurement (secondary) normal
Idc measurement (secondary) normal
Signals are not output when the limit parameter (LMT field) is OFF during
the list sweep measurement. Signals are not output when the comparator
function (COMP field) is OFF during the measurement other than list
sweep.