6-24 Maintenance
9. Use the 3 mm hex key to secure the encoder in place.
10. Remove the gauge from under the encoder.
11. Manually move the dispensing head up and down and make sure that the encoder LED
remains green when in motion.
 If the LED turns red or orange when the axis is in motion, repeat Step 4 through Step 11
and adjust the encoder height again.
 If you notice that the LED turns red or orange in a specific location, it may be due to an
obstruction or a damaged encoder strip. Try using a soft cloth and mild cleanser to clean
that location. If the problem is not resolved, repeat Step 4 through Step 11 and adjust
the encoder height again.
12. When the LED stays green throughout the entire dispensing head travel, close the
dispensing system hatch and restart Fluidmove.
13. Observe the dispensing head motion to ensure the dispensing head initializes properly and
finds home correctly.