Troubleshooting 7-9
7.4.10 Vision System
Table 7-10 Vision System Troubleshooting
Symptom
Possible Cause
Recovery
No Image
(Fluidmove display
window is a solid
pink color)
Camera not connected.
Verify camera cable is securely connected to the
connector on the top of the camera.
Faulty camera cable. Replace camera cable.
Camera LED is OFF
Camera not connected.
Verify camera cable is securely connected to the
connector on the top of the camera.
Faulty camera cable. Replace camera cable.
Camera LED
Steady Red
Camera not initialized.
It may take several minutes for the camera to acquire
a connection with the laptop. DO NOT attempt to run
Fluidmove or DALSA programs during this time.
Camera LED Slow
Flashing Red
Camera Initialization problem.
1. Shutdown the computer and turn off power to the
machine.
2. Restart the machine, and then the computer.
3. If necessary, contact Nordson ASYMTEK
Technical Support.
Camera LED Fast
Flashing Red
Camera overheated. Contact Nordson ASYMTEK Technical Support.
Camera LED Fast
Flashing Blue
Ethernet cable disconnected.
Verify Ethernet cable is securely connected to the
camera and computer.
No Image
(Fluidmove display
window is dark or
black)
Lighting intensity is too low.
Increase light level in Fluidmove. See 4.8.2 Lighting
and Color Adjustment.
Lighting module disconnected. Restore lighting module connection.
Lens is blocked by foreign matter.
1. Remove and inspect the camera lens.
2. If necessary, replace lens.
See 8.11 Replacing the Camera Lens.
Lighting module is not functioning
properly.
Contact Nordson ASYMTEK Technical Support.
Unfocused image
Substrate is not within height
focus limits of the camera or the
lens is loose.
Adjust camera bracket. See 5.5 Focusing the
Camera.
Lens is damaged or there is
foreign matter on the lens.
1. Remove and inspect the camera lens.
2. If lens is damaged or dirty, contact Nordson
ASYMTEK Technical Support.
See 8.11 Replacing the Camera Lens.
Inaccurate
dispensing after
moving or replacing
camera or lighting
module
Camera-to-needle offset is
incorrect and vision system
requires calibration.
1. Recalibrate camera.
2. Perform a “Valve Offsets” routine in Fluidmove.
For assistance, refer to the Fluidmove User
Guide or Online Help.
See 5.6 Calibrating the Camera.
No monitor display
Computer power switch OFF (0). Turn ON (I) computer power switch.
Monitor power or video cable is
not connected.
Verify that power and video cables are securely
connected to the back of the monitor.
Fluidmove vision
error encountered
during fiducial find
sequence.
Monitor power switch
OFF (0)
. Turn
ON (I)
monitor power switch.
Corrupted vision file.
Re-teach the vision targets or entire program as
necessary.
Loose camera lens. Refocus the camera. See 5.5 Focusing the Camera.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nordson Spectrum II Series and is the answer not in the manual?

Nordson Spectrum II Series Specifications

General IconGeneral
BrandNordson
ModelSpectrum II Series
CategoryDispenser
LanguageEnglish

Related product manuals