4 Patient Alarms and INOPs
104
cl NBP Serv Batt
INOP tone
The battery in the NBP Pod has reached the end of its useful life. It can no longer be charged. Contact
your service personnel to replace the battery.
cl SpO₂ Batt Empty
!!cl SpO₂ BatEmpty
!!!cl SpO₂ BtEmpty
Corresponding INOP tone
The remaining battery time of the SpO
2
Pod is below 30 minutes. Charge the battery.
cl SpO₂ Batt Incmp
INOP tone
The battery in use with the SpO
2
Pod is incompatible. Replace it with one approved for use with the
SpO
2
Pod.
cl SpO₂ Batt Low
INOP tone
The remaining battery time of the SpO
2
Pod is below 2 hours.
cl SpO₂ Batt Malf
INOP tone
There is a malfunction in the SpO
2
Pod's battery system. Contact your service personnel.
cl SpO₂ Batt Temp
INOP tone
The temperature of the battery in the SpO
2
Pod is critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove the Pod from the patient and contact your
service personnel.
cl SpO₂ Check Batt
INOP tone
The battery in the SpO
2
Pod is nearing the end of its useful life. Only 50 charge/discharge cycles
remain. Contact your service personnel to replace the battery.
cl SpO₂ Chk ChrgIF The interface between the SpO
2
Pod and the Charging Station is not working properly. Try cleaning
the contacts and reposition the Pod on the charging station.
cl SpO₂ Chk Sett The SpO
2
Pod's settings may have changed or become corrupt. Check the settings.
cl SpO₂ Disconnect
INOP tone
The SpO
2
Pod has lost the connection to the monitor.
cl SpO₂ No Cradle
INOP tone
The SpO
2
Pod is not in its cradle.
cl SpO₂ Remove
INOP tone
The temperature of the battery in the SpO
2
Pod is too high. Remove the SpO
2
Pod from the patient
and contact service personnel.
cl SpO₂ Serv Batt
INOP tone
The battery in the SpO
2
Pod has reached the end of its useful life. It can no longer be charged. Contact
your service personnel to replace the battery.
clSpO₂ License Req The SpO
2
Pod requires a valid software license for operating.
INOP Message, Indication What to do