41 Default Settings Appendix
520
SvO2 Default Settings
ScvO2 Default Settings
EEG Default Settings
ScvO
2
Settings Factory Adult Factory Pedi Factory Neo
Low Limit 70% 70% 70%
High Limit 80% 80% 80%
Alarms On On On
Light Intensity On On On
Color Yellow Yellow Yellow
SvO
2
Settings Factory Adult Factory Pedi Factory Neo
Low Limit 60% 60% 60%
High Limit 80% 80% 80%
Alarms On On On
Light Intensity On On On
Color Yellow Yellow Yellow
Hb/Hct Entry Hct [%]
Factor Entry Disabled
ScvO
2
Settings Factory Adult Factory Pedi Factory Neo
Low Limit 70% 70% 70%
High Limit 80% 80% 80%
Alarms On On On
Light Intensity On On On
Color Yellow Yellow Yellow
EEG Settings Factory Adult Factory Pedi Factory Neo
TP, SEF OnOnOn
MDF, PPF, Delta, Theta, Alpha, Beta Off Off Off
SEF Threshold 90% 90% 90%
Numeric Average 8 sec 8 sec 8 sec
Wave Scale 100µV (or ± 50µV if Show Gridlines is configured to Yes)
Show Gridlines No No No
Low Filter 0.5 Hz 0.5 Hz 0.5 Hz
High Filter 30 Hz 30 Hz 30 Hz
Impedance Limit 5 kOhm 5 kOhm 5 kOhm
Smoothing CSA On On On