5 Managing Patients and Equipment
114
5
At the new location, connect the MMS to the monitor. If the monitor detects a patient mismatch,
a window will open showing your patient's data and asking whether to
Complete transfer of this
patient?
.
6 Select Yes to complete the transfer. This re-admits the patient from the transfer list to the new
monitor. This will upload the patient demographics, and, if configured, the measurement settings
and trend data stored in the MMS to the receiving monitor.
7 Verify that the settings for patient category and paced mode are correct.
WARNING
Measurements from an MMS extension connected to an X2 are not available when the X2 is running
on its own battery power. They are only available when the X2 is powered by external power: when
connected to a host monitor, to the external power supply (M8023A) or to the Battery Extension
(865297).
Transferring a Centrally-Monitored Patient with X2
PIIC Scenario: A centrally-monitored patient is moved to another monitoring location on the same
Information Center database server without interrupting the collection of patient trend information.
1 Before you disconnect the X2 from the host monitor, select the patient name field or select the
Patient Demogr. SmartKey to open the Patient Demographics window, then select the Transfer
pop-up key. If the patient is not admitted or not monitored by an Information Center, the
Transfer key is inactive ("grayed-out").
This step preserves the patient's demographic data during the transfer.
2 Remove the X2 and any connected extensions from the monitor.
3 Silence the resulting MMS Unplugged INOP.
4 Move the patient using the X2 as the transport monitor.
5 At the new location, connect the X2 to the new host monitor. If the monitor detects a patient
mismatch, a window will open showing your patient's data and asking
Complete transfer of this
patient?
.
6 Select Yes to complete the transfer. This re-admits the patient from the transfer list to the new
monitor. This will upload the patient demographics, and, if configured, the measurement settings
and trend data stored in the X2 to the receiving monitor.
7 Verify that the settings for patient category and paced mode are correct.
Transferring a Centrally-Monitored Patient with X2 using IIT
PIIC Scenario: A centrally-monitored patient is moved with an X2 to another monitoring location on the
same Information Center database server without interrupting the collection of patient trend
information. The X2 has an IntelliVue Instrument Telemetry interface (IIT), is connected to a host
monitor, and is declared as a telemetry device at the Information Center.
NOTE
The Transfer key is not available while the X2 is connected to a host monitor (Companion Mode is
indicated).
1 Remove the X2 and any connected extensions from the host monitor.