EasyManua.ls Logo

Philips MX400 - Adjusting the Calibration Factor; Displaying a Mean Pressure Value Only; Changing the Pressure Wave Scale; Optimizing the Waveform

Philips MX400
534 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12 Monitoring Invasive Pressure
217
Adjusting the Calibration Factor
Each time you use a reusable transducer, compare the calibration factor written on your transducer
with the calibration factor shown on the monitor. To ensure accurate measurement, they must be the
same.
1 In the Setup <Press Label> menu, select Cal. Factor.
If the value here does not match that on the transducer, select the corresponding value from the
list now in accordance with your hospital's procedure.
2 To confirm you want to use the new calibration factor, select the Confirm pop-up key.
Displaying a Mean Pressure Value Only
Use this when you want to see only the mean pressure.
In the pressure's setup menu, select
Mean only. Toggle between On to display mean pressure value
only, and
Off to display all pressure values (systolic, diastolic and mean).
Changing the Pressure Wave Scale
1 Select the label of the pressure wave whose scale you want to set to enter the Setup <Press
Label>
menu.
2 In the Setup <Press Label> menu, (for example ABP) select Scale.
3 Select a value from the pop-up list:
a positive value sets the top gridline. The bottom gridline is set at zero.
a negative value sets the bottom gridline. The middle gridline is set at zero.
Optimizing the Waveform
In the Setup <Press Label> menu, select Optimum Scale to let the monitor select the best minimum
and maximum scales for the current wave.
Using the Wave Cursor
When the pressure wave is selected, you can display a cursor in the wave. It appears as a white
horizontal line in the wave. Using the pop-up keys you can move the cursor up or down to the
required position and store the corresponding value. The cursor value can be stored
as a systolic, diastolic or mean pressure value,
as an IAP (intra-abdominal pressure) value - for the waves P, P1 to P8, if IAP is configured as a
measurement which can be manually entered
as a PAWP value - if the pressure wave is a PAP wave
The stored value appears in the trend database as a manually entered value.
To display and position the cursor
1 Select the pressure wave.
2 Select Activate Cursor.
3 Use the pop-up arrow keys to position the cursor.

Table of Contents

Related product manuals