5 Managing Patients and Equipment
125
The ECG measurement will be activated again at the monitor. [Note that in this case, as the screen
switches back to the monitor's own measurements, the SpO
2
T measurement (if present) will no longer
be displayed].
In the same way the source is tracked when a telemetry device is directly connected to a monitor, then
disconnected and vice versa.
In case of ambiguity, a yellow INOP message
!!Check ECG Source indicates that more than one valid
ECG source is active.
Synchronized Settings
For some measurements, settings can be synchronized between the monitor and another measurement
device. For example, if ECG is measured at the monitor, and then the patient is connected to a
telemetry device for monitoring, the Information Center will use the monitor settings for the telemetry
device. In general, the following settings will be synchronized:
PIIC iX With the IntelliVue Information Center iX, the following additional measurement settings can be
synchronized:
Heart Rate HR/Pulse Alarm On/Off, Heart Rate High/Low Limit,
ECG ECG On/Off
1
, Primary Lead, Secondary Lead, Va Lead
2
, Vb Lead
2
, Lead
Placement
Arrhythmia Analysis Mode, Arrhythmia On/Off, Asystole Threshold, Pause Threshold,
VTach HR, VTach Run, PVCs/min, Vent. Rhythm, SVT HR, SVT Run,
PVCs/min On/Off, Pacer not capture On/Off, Pacer not pace On/Off,
Non-Sustain On/Off, Vent. Rhythm On/Off, Run PVCs On/Off, Pair
PVCs On/Off, Missed Beat On/Off, Pause On/Off, R-on-T PVCs On/
Off, Vent. Bigeminy On/Off, Vent. Trigeminy On/Off, Multiform PVCs
On/Off, Irregular HR On/Off, SVT On/Off, Afib On/Off, Afib/IrrHR
End Threshold, All ECG Alarm INOP mode.
ST ST Analysis On/Off, ST Alarm On/Off, ISO Point, J point, ST point, ST
Alarm Limits, ST-Index On/Off, ISO/J-Point Detection, ST Lead On/
Off, ST Baseline
QT QT analysis On/Off, QT Lead, QTc High Alarm On/Off, ΔQTc Alarm
On/Off, QTc High Limit, ΔQTc High Limit, QT Baseline
3
SpO
2
TSpO
2
Alarms on/off, SpO
2
Alarm limits, SpO
2
Low Alarm Delay, SpO
2
High Alarm Limit, Desat Alarm Limit
NBP Alarm Suppression On/Off, Pulse(SpO
2
) On/Off
1
ECG On/Off setting is forced to ON by the monitor, if the local setting is ON.
2
Va and Vb leads are reset to default (V2, V5) if the configured Va or Vb lead for the telemetry device is not one of V1
through V6.
3
With PIIC, QT baseline is synchronized, but not QT snippets.
ECG/Arrhythmia Some ECG Alarms INOP On/Off
ST STE On/Off, STE Alarm On/Off
NBP Sys/Dia/Mean Alarm Limits, Alarms On/ Off, Alarm Source
Resp Apnea Time, Alarm Limits, Alarm On/Off, Resp On/Off