40 Specifications
482
NBP
Complies with IEC 60601-2-30:1999/EN 60601-2-30:2000.
9% 10 sec 10 sec 11 sec
>9% 10 sec 10 sec 10 sec
SpO
2
Smart Alarm Delay - Detailed Specifications
Deviation from violated
alarm limit
Resulting alarm delay according to selected mode
Short Medium Long
NBP Performance Specifications
Measurement Ranges Systolic Adult: 30 to 270 mmHg (4 to 36 kPa)
Pedi: 30 to 180 mmHg (4 to 24 kPa)
Neo: 30 to 130 mmHg (4 to 17 kPa)
Diastolic Adult: 10 to 245 mmHg (1.5 to 32 kPa)
Pedi: 10 to 150 mmHg (1.5 to 20 kPa)
Neo: 10 to 100 mmHg (1.5 to 13 kPa)
Mean Adult: 20 to 255 mmHg (2.5 to 34 kPa)
Pedi: 20 to 160 mmHg (2.5 to 21 kPa)
Neo: 20 to 120 mmHg (2.5 to 16 kPa)
Pulse Rate Adult: 40 to 300 bpm
Pedi: 40 to 300 bpm
Neo: 40 to 300 bpm
Accuracy Max. Std. Deviation: 8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
Pulse Rate Measurement Accuracy 40 to 100 bpm: ±5 bpm
101 to 200 bpm: ±5% of reading
201 to 300 bpm: ±10% of reading
(average over NBP measurement cycle)
Pulse Rate Range 40 to 300 bpm
Measurement Time Typical at HR >60 bpm
Auto/manual/sequence:
Adult: 30 seconds
Neonatal: 25 seconds
Stat: 20 seconds
Maximum time:
Adult/pedi: 180 seconds
Neo: 90 seconds
Cuff Inflation Time Typical for normal adult cuff: Less than 10 seconds
Typical for neonatal cuff: Less than 2 seconds
Initial Cuff Inflation Pressure Adult: 165 ±15 mmHg
Pedi: 130 ±15 mmHg
Neo: 100 ±15 mmHg
Maximum Cuff Pressure Adult/Pedi: 300 mmHg
Neo: 150 mmHg
Auto Mode Repetition Times 1 min, 2 min, 2.5 min, 3 min, 10 min, 15 min, 20 min, 30 min, 45 min, 1h, 2h,
4h, 8h, 12h, 24h
STAT Mode Cycle Time 5minutes