EasyManua.ls Logo

Philips MX400 - Starting the Different Types of Report

Philips MX400
534 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29 Printing Patient Reports
370
When an X2 is connected to a host monitor, reports requested or stored on the X2 can be printed via
the host monitor. Reports requested on the X2 when connected to a host monitor will not show any
patient alarms or INOPs. In order to see alarms, reports should be requested on the host monitor.
Starting the Different Types of Report
Report types Report contents How to start printing reports
Vital Signs Report Depends on selected trend group, data
resolution, and period.
In the Vital Signs window, select Print/ Record, then
select Print.
Graphic Trends Report In the Graphical Trends window, select Print.
Events Reports Event Episode In the Event Episode or Event Review window, select
the pop-up key Print/ Record, then select Print
Review or Print Episode.
Graphic or tabular Event Review
EEG Report EEG Waves, numerics, CSAs, montage
information and impedance values
In the Setup EEG menu, select Print Report.
ECG Reports Depends on format selected. Select the ECG Report A or ECG Report B SmartKeys,
if configured, or the Print Reports SmartKey followed
by ECG Report A or ECG Report B.
12-Lead Report Captured ECG waves, measurement values, ST
Map (optional) and analysis results from the
Information Center (PIIC iX only)
In an active 12-Lead application window, select the Print
Report pop-up key
ST Map Report ST data in a transversal and/or horizontal multi-
axis diagram,
In the ST Map (Current) or ST Map (Trend) window,
select Print.
Cardiac Output Report Trial curves and numerics In the Cardiac Output (Right Heart) or Cardiac
Output (Transpulmonary) window, select the pop-up
key Print/ Record.
Wedge Report Wedge numerics and reference wave In the Wedge window, select the Print Wedge pop-up
key.
Calculations Report Hemodynamic, Oxygenation, or Ventilation
Review
In the Calculations windows, select the Print/ Record
pop-up key.
Drug Calculator Report Titration Table In the Titration Table window, select the Print
Titr.Tbl pop-up key.
Drip Table In the Drip Table
window, select the Print Drip Tbl
pop-up key.
Alarm Limits Report Graphic and numeric report of all current alarm
limits
In the Alarm Limits window, select Print Limits.
Realtime Reports,
including oxyCRG
Reports
Patient data and numerics, and either:
all displayed waves OR all measured waves OR
all measured RT waves OR all measured HiRes
waves, OR oxyCRG waves (RT waves: ECG
Primary lead, Pleth, Resp; HiRes waves: btbHR,
HiRes SpO
2
, HiRes Resp)
Select the Realtime Report SmartKey, if configured.
Loops Report Individual Loops diagram with timestamp,
currently measured numerics
In the Loops window, select Print Loop.
Central Report A/B/C These are reports configured at the Information
Center. Monitors connected via the IntelliVue
Instrument Telemetry wireless network can have
these reports available.
In the Main Setup select Reports then Central
Report A, Central Report B or Central Report C.

Table of Contents