40 Specifications
492
tcGas
Complies with IEC 60601-2-23:1999/EN 60601-2-23:2000.
CCO Range 0.1 to 25.0 l/min
Standard Deviation 10% or 0.3 l/min
Display Update 2 seconds nominal
Blood Temperature Range 17.0 to 43°C (62.6 to 109,5°F)
Injectate Temperature Range -1 to 30°C (30.2 to 86.0°F)
C.O./CCO Performance Specifications
C.O./CCO Alarm
Specifications
Range Adjustment Delay
TBlood 17 to 43°C Steps of 0.5°C (17 to 35°C)
Steps of 0.1°C (35 to 43°C)
Steps of 1°F (63 to 95°F)
Steps of 0.2°F (95 to 109°F)
10 seconds after the value exceeds
the set limit range
CCO 0.1 to 25.0 l/min 0.1 l/min (0.1 to 10.0 l/min)
0.5 l/min (10.0 to 25.0 l/min)
10 seconds after the value exceeds
the set limit range
tcGas Performance Specifications
tcpO2 Range 0 to 750 mmHg (0 to 100 kPa)
Resolution 1 mmHg (0.1 kPa)
Overall Accuracy
(incl. transducer)
0 to 20.9% O
2
better than +5 mmHg (+0.7 kPa)
Above 20.9% O
2
better than +10% of reading
tcpO2 Test Signal 60 mmHg (8.0 kPa)
tcpO2 Drift, including transducer Typically 1.25%/h
tcpO2 response time, including transducer Typically 30 s
tcpCO2 Range 5 to 200 mmHg (0.7 to 26.7 kPa)
Resolution 1 mmHg (0.1 kPa)
Overall Accuracy
(incl. transducer)
5 to 76 mmHg (0.7 to 10.1 kPa) better than +5 mmHg (+0.7 kPa)
Above 76 mmHg (10.1 kPa) better than +10% of reading
tcpCO2 Test Signal 40 mmHg (5.3 kPa)
tcpCO2 Drift, including transducer Typically 2.5%/h
tcpCO2 response time, including transducer Typically 60 s
Warm-up Time <3 minutes
Site Timer (M1018A
only)
0.5, 1, 1.5, 2, 2.5, 3, 4, 5, 6, 7, or 8 hours. Change Site alarm when site time is
expired and configurable automatic heating switch-off
Transducer Heating Available Temperatures -
37.0°C, 41.0 to 45°C in steps of 0.5°C