41 Default Settings Appendix
510
Arrhythmia Default Settings
Asystole Thresh. 4.0 sec 4.0 sec 3.0 sec
Δ ExtrTachy 20 bpm 20 bpm 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm
Δ ExtrBrady 20 bpm 20 bpm 20 bpm
Brady Clamp 40 bpm 40 bpm 50 bpm 40 bpm 60 bpm 80 bpm
ECG Alm Off INOP Cyan Cyan Cyan
Fallback On On On
Alarms Off Enabled Enabled Enabled
AlarmSource Sel. Enabled Enabled Enabled
Va Lead V2 V2 V2
Vb Lead V5 V5 V5
SyncPulse Sensit Medium Medium Medium
SyncPulse Marker On On On
PulseAlarms Tele Enabled Enabled Enabled
ECG Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Arrhythmia Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Arrhythmia On On Off Off Off Off
Pause Threshold 2.0 sec 2.0 sec 1.5 sec
VTach HR 100 120 150
VTach Run 555
Vent Rhythm 14 14 14
SVT HR 180 200 210
SVT Run 555
PVCs/min 10 5 5
Non-Sustain On On On
Vent Rhythm On On On
Run PVCs On On On
Pair PVCs On On On Off Off Off
R-on-T PVCs On On On Off Off Off
V.Bigeminy On On On Off Off Off
V.Trigeminy On OnOnOffOffOff
PVCs/min On On On Off Off Off
Multif.PVCs On OnOnOffOffOff
Pacer n.Cap On On On
Pacer n.Pac On On On
Pause On OnOnOffOffOff
Missed Beat On On On Off Off Off