41 Default Settings Appendix
515
Invasive Pressure Default Settings
ABP, P, P1, P2, P3, P4,
UAP, ART, Ao, FAP, BAP
Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Alarms from Sys. Sys. Sys.
High Limit 160/90 (110) 120/70 (90) 90/60 (70) 180/90 (110)
Low Limit 90/50 (70) 70/40 (50) 55/20 (36) 70/50 (70)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High 15 10 5
Δ Extreme Low 15 10 5
Sys. High Clamp/Dia.
High Clamp (Mean
High Clamp)
190/100 (125) 140/80 (100) 105/75 (75)
Sys. Low Clamp/Dia.
Low Clamp (Mean Low
Clamp)
80/45 (65) 60/35 (45) 45/15 (30) 65/45 (65)
Scale 150 100 100
Mean only No No No
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr. 60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Red Red Red
Pulse settings for all arterial pressures except P1 to P4
Pulse (<Press Label>) On On On
For all other Pulse default settings, refer to the Pulse Default Settings table.
CVP, RAP, LAP, UVP
Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Alarms from Mean Mean Mean
High Limit 14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High 555
Δ Extreme Low 555
Sys. High Clamp/ Dia.
High Clamp (Mean
High Clamp)
20/10 (15) 15/5 (10) 15/5 (10)
Sys. Low Clamp/Dia.
Low Clamp (Mean Low
Clamp)
0/-5 (-5) 0/-5 (-5) 0/-5 (-5)
Scale 30 30 30