41 Default Settings Appendix
516
Mean only Yes Yes Yes
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr. 60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Cyan (light
blue)
Cyan (light
blue)
Cyan (light
blue)
Blue Blue Blue
CVP, RAP, LAP, UVP
Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
PAP Settings
Factory Defaults
Adult Pedi Neo
Alarms from Dia. Dia. Dia.
High Limit 34/16 (20) 60/4 (26) 60/4 (26)
Low Limit 10/0 (0) 24/-4 (12) 24/-4 (12)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High 555
Δ Extreme Low 555
Sys. High Clamp/Dia. High Clamp (Mean
High Clamp)
45/20 (25) 65/5 (35) 65/5 (35)
Sys. Low Clamp/Dia. Low Clamp (Mean
Low Clamp)
5/-5 (-5) 15/-5 (5) 15/-5 (5)
Scale 30 30 30
Mean only No No No
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr. 60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Yellow Yellow Yellow
ICP, IC1, IC2 Settings
Factory Defaults
Adult Pedi Neo
Alarms from Mean Mean Mean
High Limit 14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High 10 10 10
Δ Extreme Low 10 10 10
Sys. High Clamp/Dia. High Clamp (Mean
High Clamp)
20/10 (15) 15/5 (10) 15/5 (10)