4 Patient Alarms and INOPs
90
SpO2 INOPs
Tven INOPs See <Temp Label> INOPs
Tvesic INOPs See <Temp Label> INOPs
INOP Message, Indication What to do
INOP Message, Indication What to do
<SpO Label> Chk Sensor The condition of the signal at the SpO
2
sensor is not as expected. Accuracy may be compromised.
Check proper application of the sensor in accordance with the Instructions for Use. If the INOP
persists, consider exchanging the sensor.
<SpO Label> Deactivated
INOP tone
The SpO
2
measurement label in the measurement device has been deactivated by deactivating the label
in the Measurement Selection window. The measurement automatically disappears from the
display. To switch the measurement on again, reactivate the measurement label in the Measurement
Selection window.
<SpO Label> Equip Malf
Numeric is replaced by -?-
INOP tone
The MMS or module is faulty. Unplug and replug the MMS or module. If the INOP persists, contact
your service personnel.
<SpO Label> Erratic
Numeric is replaced by -?-
INOP tone
Check the sensor placement. Try another adapter cable and sensor. If the INOP persists, contact your
service personnel.
<SpO Label>
Extd.Update
Numeric is replaced by -?-
(questionable numeric)
The update period of displayed values is extended due to an NBP measurement on the same limb or
an excessively noisy signal.
<SpO Label>Interference
Numeric is replaced by -?-
INOP tone
There is too much interference, caused by a high level of ambient light and/or electrical interference.
Cover the sensor to minimize ambient light. If the INOP persists, make sure that the sensor cable is
not damaged or positioned too close to power cables.
<SpO Label> Low Perf
Numeric is replaced by -?-
(questionable numeric)
Accuracy may be compromised due to very low perfusion. Stimulate circulation at sensor site. If INOP
persists, change the measurement site.
<SpO Label> No Pulse
Numeric is replaced by -?-
INOP tone
Check the perfusion at measurement site. If necessary, stimulate circulation or change measurement
site. If the INOP is due to NBP measurement on the same limb, wait until the NBP measurement is
finished.
<SpO Label> No Sensor
Numeric is replaced by -?-
INOP tone
Make sure the SpO
2
sensor is connected. If the INOP persists, try another adapter cable and sensor. If
you silence this INOP, the measurement will be switched off.
<SpO Label> NoisySignal
Numeric is replaced by -?-
INOP tone
Excessive patient movement or electrical interference is causing irregular pulse patterns. Try to reduce
patient movement or to relieve the cable strain on the sensor.
<SpO Label> Poor Signal
Numeric is replaced by -?-
(questionable numeric)
The signal condition of the SpO
2
measurement is poor and measurement accuracy may be
compromised. If the INOP persists, consider changing the application site or using another sensor.
<SpO Label> Pulse?
Numeric is replaced by -?-
INOP tone
The detectable pulsations of the SpO
2
signal are outside the specified pulse rate range.
<SpO Label> Searching
Numeric unavailable
SpO
2
is analyzing the patient signal to derive Pulse, SpO
2
and Perf values. Please wait until the search
analysis is complete.
<SpO Label> Sensor Malf
Numeric is replaced by -?-
INOP tone
The SpO
2
sensor or adapter cable is faulty. Try another adapter cable and sensor. If the INOP persists,
contact your service personnel.